Wie wird das zu testende Gerät (Zelle, Modul oder Pack) mit dem Arbin-System verbunden?

Mai 8, 2026

There are several way to connect to the DUT as follows: Individual cables Individual cell holders (coin cells, cylindrical cells, flat/pouch cells) Trays for holding multiple cells (coin cells, cylindrical cells, flat/pouch cells) Racks for organizing cells Chambers We will go to each method one by one as follows: 1. Individual cables It could be […]

Meine Prüfung brach wegen eines Fehlers mit der Bezeichnung “IR Inf” ab.”

Mai 6, 2026

The “IR Inf” error occurs when the system fails to obtain a valid result during an Internal Resistance (IR) measurement step. This typically happens when the measurement conditions are not sufficient to produce a reliable calculation, causing the system to return an infinite (invalid) resistance value and stop the test. Solution: Make sure the amplitude […]

Kann ich eine MTCI-Einheit kaufen und zwischen den Kammern wechseln?

Mai 6, 2026

The answer is yes and no. Yes, if those chambers use the same communication protocol. For example, Votsch VT4002 and Weiss WT20/40 use the Votsch L4006 protocol. If you have these two chambers, you only need one MTCI unit. However, if you have another chamber like Espec SU-241 that uses the Espec SCP220/SCP-N300/SCP310 protocol, you […]

Kann die Arbin-Software einen Graphen erstellen, der Daten über Zyklen für einen oder mehrere Kanäle hinweg vergleicht?

Mai 6, 2026

Yes. Arbin Data Watcher software (Version 2.2.217 and later) can do this. In Data Watcher software, click “Edit Chart Template”. From the Model dropdown, select “CycleCompare”. Return to the main window of Data Watcher. In the axis configuration, you can select the physical quantity for the X, Y1, and Y2 axes. In this example, X […]

Schrittsteuerungsfehlerprüfung unsicher

Mai 6, 2026

The “Step Control Error Check Unsafe” warning indicates that the system has detected a significant deviation between the expected (configured) output and the actual measured output during a test step. They are designed to detect open circuits or short circuits caused by loose IV cable connections, it also detects abnormal output when the system has […]

Mein Lade-/Entladestrom ist fast 0 oder viel niedriger als der eingestellte Wert?

Mai 3, 2026

In some cases, the charge or discharge current may be significantly lower than the configured value, or even close to zero. This issue is typically caused by software configuration limits or hardware-related problems. This article provides a structured troubleshooting guide to help identify and resolve the issue efficiently. Symptoms You may observe the following: The […]

Mein Test wurde aufgrund eines Fehlers namens “Tokenbroken” abgebrochen.”

April 9, 2026

This error indicates too much data logged at a very short period of time. Check the schedule, make sure the data log setting is appropriate. Please open the schedule and locate the Log Limit: Then check the following points: Logging interval (time between data points)Many Arbin machines can acquire data at rates of up to […]

Warum sehe ich den Hilfswert nicht, obwohl ich ihn bereits zugeordnet habe?

April 5, 2026

In some cases, users may successfully configure (map) an auxiliary channel, but the corresponding auxiliary values do not appear during test execution. This issue is typically caused by missing schedule assignment or incomplete auxiliary channel configuration within the schedule. Possible Causes No schedule or an incorrect schedule is assigned to the channel Forgot to add […]

Warum stoppt mein Zeitplan sofort nach dem Start?

April 2, 2026

In some situations, a schedule may stop immediately after it starts. This behavior usually occurs when certain safety conditions are not met or when required devices are not properly connected. The system automatically stops the schedule to prevent unexpected behavior, hardware damage, or unsafe operating conditions. This article provides common causes and troubleshooting steps to […]